120 SHARP TOP-BOTTOM

1. Test overview.

This test is used to adjust the film carrier so that the sharpness at the top of film is as close as possible to sharpness in bottom of film.
For this, the scanner will scan a target film and will determine a level of sharpness in top and bottom of film. The user will be required to adjust the screws of the film carrier to obtain a acceptable level of sharpness.
If the sharpness at the top is different from the sharpness in bottom of film, this can cause differences in the sharpness of the scanned images.

2. Conditions of the test.

This test relates to the 120 film carrier inserted in the scanner.
The 120 film carrier must be inserted in the scanner to carry out this test.
The film 120 Colour Target Film must be used for this test (ref 070 100 417A) :
120 Color target film
The acceptance of the test is automatic.
No parameter is saved.

3. Pre-requirements.

4. Test realisation.

Click on RUN to execute the test. A message will appear requiring the insertion of the 120 Colour target film. When the film is inserted (watch for the direction of the film), click on OK.
Note : Colour target film consists of : The scanner will scan the first and the second exposure of film then will display an image mixing :

The calculated level of sharpness is displayed at the top in the boxes TOP and BOTTOM. In the box Min to reach enter the value hand written on the film (about 70, according to films). .

Note : the sharpness value can vary from one scanning to another, sometimes of 4 or 5 points. This is due to the the fact that the film is never scanned exactly at the same place.

5. Test acceptance.

The test is accepted if the difference between the level of sharpness calculated at the top and that calculated at the bottom is lower to 6% of largest of the 2.

6. Other affected tests.

If the sharpness at the top is very different of the sharpness at the bottom, test 120 Cal encoder 2 could be affected.

7. Trouble shooting.

  • If The test is rejected with the message " Top is sharper than bottom" (or the reverse) : the level of sharpness at the top is higher or lower to those of the bottom. Withdraw the film carrier and adjust the high or low screw (while following the screen instructions). The goal is to increase the lowest value. Caution: do not turn more than 1/4 of turn in one direction or in the other in relation to the starting position.
  • If The test is rejected with the message " Nil value for sharp max " : this means that the sharpness value to be reached (Min to reach) was not indicated (it remained to 0). Start again the test and enter the value Min to reach.